您的位置:首页 > 其它

体外诊断(POCT)之检测线配置文件修改

2016-06-14 18:12 190 查看
检测线配置文件由入口配置文件的活动配置节指定, 参见入口配置文件修改

配置节由条纹配置文件指定, 参见条纹配置文件修改

以基础版本为例, 检测线配置文件的内容如下:

; ====================================================================================================
; filename is selected by application
; ====================================================================================================
Author = Valiant
Copyright = cto@renshenguo.com
Linkage = http://blog.csdn.net/prsniper Time = [2016-05-25 16:24:17]

; ====================================================================================================
; analyte configurations(select after finish parsing stripe.ini)
; ====================================================================================================
[Dummy.Stripe1.Ctrl]
Offset=4.0 ; line probably offset-start(relative to stripe)
Range=3.5 ; line probably offset-width(relative to stripe)
Type=0 ; line type: control=ignore remaining lines
Curves={0,0,0,0,0,0}
Mult=0
Caption=[n/a]
Digits=3

[Dummy.Stripe1.Test]
Offset=10.0 ; line probably offset-start(relative to stripe)
Range=3.5 ; line probably offset-width(relative to stripe)
Type=1
Curves={0,0,0,0,0,0} ; test line curves(override barcode/matrix code)
Mult=0 ; test line result multiplier(curve override if not 0)
Caption=[ng/ml] ; test line result unit
Digits=3 ; test line significant figure(0 represent using primitive)


主要的配置项目如下:
Offset为检测线水平偏移开始,大致位置,以毫米为单位

Range为水平大致宽度, 以毫米为单位

Type是检测线类型,0为控制线(C线),1为检测线(T线)

Curves为曲线参数,为测试线(T线专用),Mult为结果倍率,若为零,则包括曲线参数在内的参数有阵列码包含的内容决定

Caption是单位,比如ng/ml(纳克每毫升), 显示在软件界面上

Digits是结果保留的小数位数,影响显示界面的结果小数位数

备注:

范围是大致范围, 软件为智能的搜索范围内最明显的线条, 当然仅限给出的范围内, 设定范围是为了节省不必要的计算时间

得到线条位置后, 软件会搜索强度最好的部分进行计算, 边沿和两端的过渡区域将被舍弃, 强度值为加权均值, 消除噪点干扰

更深层次的逻辑有技术二次开发权限的用户可以获得有关文档.

检测线范围示意图:

内容来自用户分享和网络整理,不保证内容的准确性,如有侵权内容,可联系管理员处理 点击这里给我发消息
标签: